|description=[[Image:Pos_holder.JPG|right|180px]]The '''OroboPOS-Holder''' (POS-Holder), made from blue POM, is screwed into the copper block of the [[O2k-Main Unit]], guiding the [[OroboPOS-Connector]] with the [[OroboPOS]] sensor head (POS) to the 2-mL [[O2k-chamber]], and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-chamber with the [[OroboPOS-Seal Tip]]. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
|description=[[Image:Pos_holder.JPG|right|180px]]The '''POS-Holder''', made from blue POM, is screwed into the copper block of the [[O2k-Main Unit]], guiding theguiding the [[OroboPOS|POS]] to the 2-mL [[O2k-chamber]], and keeping the SmartPOS/OroboPOS-Connector in a fixed position for sealing the O2k-chamber with the [[POS-Seal Tip]]. In addition, the POS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.


Two units of this item are standard components mounted on the [[O2k-Main Unit]].
Two units of this item are standard components mounted on the [[O2k-Main Unit]].
Line 14: Line 14:


::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
== A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber ==
== A damaged POS-Holder may cause medium to leak from the O2k-Chamber ==


:::* The OroboPOS-Holder may be damaged:
:::* The POS-Holder may be damaged:


::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below'').   
::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the POS-Holder is fully inserted, and subsequently screwing tightly the POS-Holder (''see center picture below'').   
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].
::::# After fully unscrewing the POS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].


::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see'' pictures below).
::::* We recommend to check before any reassembly the mechanical integrity of the POS-Holder (''see'' pictures below).


[[File:OroboPOS-Holder002.jpg|left|200px]]
[[File:OroboPOS-Holder002.jpg|left|200px]]
Line 27: Line 27:
[[File:OroboPOS-Holder004.jpg|centre|200px]]
[[File:OroboPOS-Holder004.jpg|centre|200px]]


* Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
* Left picture: brand new POS-Holder, no damage on the thread, no damage on the edge.
* Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.  
* Center picture: used POS-Holder, almost no damage on the thread, the edge is slightly damaged.  
* Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
* Right picture: heavily used POS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.




::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables]
::::* Replacement of a damaged POS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables]
  ''See'' [[Talk:OroboPOS-Holder|Discussion page]]
  ''See'' [[Talk:POS-Holder|Discussion page]]


== Keywords ==
== Keywords ==

Latest revision as of 11:13, 27 July 2023


                  


O2k-Open Support

POS-Holder



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Pos holder.JPG

The POS-Holder, made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding theguiding the POS to the 2-mL O2k-chamber, and keeping the SmartPOS/OroboPOS-Connector in a fixed position for sealing the O2k-chamber with the POS-Seal Tip. In addition, the POS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.

Two units of this item are standard components mounted on the O2k-Main Unit.


Reference: Product details and purchase information

Assembly

A damaged POS-Holder may cause medium to leak from the O2k-Chamber

  • The POS-Holder may be damaged:
  1. During chamber assembly, when fixing the chamber in a wrong rotational position with the O2k-Chamber Holder (from above) before the POS-Holder is fully inserted, and subsequently screwing tightly the POS-Holder (see center picture below).
  2. After fully unscrewing the POS-Holder and forcing during reassembly an improperly aligned holder into the thread of the O2k-Main Unit.
  • We recommend to check before any reassembly the mechanical integrity of the POS-Holder (see pictures below).
OroboPOS-Holder002.jpg
OroboPOS-Holder003.jpg
OroboPOS-Holder004.jpg
  • Left picture: brand new POS-Holder, no damage on the thread, no damage on the edge.
  • Center picture: used POS-Holder, almost no damage on the thread, the edge is slightly damaged.
  • Right picture: heavily used POS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.


See Discussion page