Difference between revisions of "O2k-OpenSupportAlert2020 Archive"
From Bioblast
Line 8: | Line 8: | ||
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]] | | width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]] | ||
|'''Have you observed an 'O<sub>2</sub> slope neg.' > 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup> in the closed chamber without a biological sample? '''<br> | |'''Have you observed an 'O<sub>2</sub> slope neg.' > 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup> in the closed chamber without a biological sample? '''<br> | ||
This may be due to [ | This may be due to [[biological contamination]] of the medium or the O2k-Chamber. <br> | ||
For further details on how to clean the O2k-Chamber properly, see: <br> | For further details on how to clean the O2k-Chamber properly, see: <br> | ||
- » [ | - » [[O2k-Chamber_MitoPedia#Cleaning_the_O2k-Chamber|Cleaning of the O2k-Chamber]]«<br> | ||
|} | |} | ||
---- | ---- |
Revision as of 08:27, 18 August 2020
- Currently Answered Questions from the »Oroboros Ecosystem Agenda«
Aug 2020
- Aug-17 Mo
Have you observed an 'O2 slope neg.' > 4 pmol∙s−1∙mL−1 in the closed chamber without a biological sample?
This may be due to biological contamination of the medium or the O2k-Chamber.
For further details on how to clean the O2k-Chamber properly, see:
- » Cleaning of the O2k-Chamber«
- Aug-03 Mo
What do you do if values, symbols, and/or units are not correctly displayed in the DatLab software?
It is possible to retrieve the values, symbols and/or correct units by updating the Windows control panel settings.
- » Troubleshooting for problems with text in DatLab«