The illumination in both chambers is switched on/off.
Abbreviation: F10
DatLab 8
Last update: 2024-05-06
- Illumination on/off: The illumination in both chambers is switched on/off. When on, the illumination is automatically switched off after 60 seconds to avoid interference with optical
measurements. In fact, this illumination must be distinguished from light introduced into the chambers by LEDs for the purpose of spectrophotometric and fluorometric measurements and from the PBLS in the PhotoBiology-Module. For these applications, the internal illumination must be switched off.
DatLab 7
- Illumination on/off: The illumination in both chambers is switched on/off.
MitoPedia O2k and high-resolution respirometry: