* From [[MiPNet06.03 POS-Calibration-SOP|MiPNet06.03 POS-Calibration-SOP (Version 14)]]: Prepare "zero solution" by dissolving c. 20 mg sodium hydrosulfite ([[OroboPOS-Service Kit]]) or two tips of a spatula in 0.5 ml of water. Mix in a small vial with minimum gas space. Use fresh. Dithionite may not work after prolonged storage. |
Revision as of 16:26, 31 August 2015
Description
Zero oxygen solution powder, Na2S2O4, used for calibration of oxygen sensors at zero oxygen, or for stepwise reduction of oxygen concentration in instrumental O2 background tests.
Abbreviation: Na2S2O4
Reference: MiPNet06.03 POS-Calibration-SOP, MiPNet14.06 InstrumentalO2Background
MitoPedia methods:
Respirometry
Further inforation >> Talk:Dithionite
- From MiPNet06.03 POS-Calibration-SOP (Version 14): Prepare "zero solution" by dissolving c. 20 mg sodium hydrosulfite (OroboPOS-Service Kit) or two tips of a spatula in 0.5 ml of water. Mix in a small vial with minimum gas space. Use fresh. Dithionite may not work after prolonged storage.