Air calibration for high O2 experiments

From Bioblast
Revision as of 09:42, 2 September 2011 by Fasching Mario (talk | contribs) (created page based on customer question)
(diff) ← Older revision | Latest revision (diff) | Newer revision β†’ (diff)

Question: For a experiment with with high (above atmospheric levels) O2 concentrations I have to change the gain and therefore to perform a new "air calibration". How to do this?

Answer: The air calibration has always to be done at atmospheric saturation (a point of known O2 conc.), but at the correct Gain

  1. set the gain to 1
  2. perform air calibration at atmospheric saturation, open chamber, near zero flux at open chamber, as usually
  3. close the chamber
  4. wait until you observe a stable flux at closed chamber near to air saturation of 2 to 4 pmol/s ml (quality control for biological contamination)
  5. increase the O2 concentration either with a gas phase method or (in MiR06) with H2O2 injections.


See also: Air calibration, Raw_Signal, Gain (O2_channel)



Template:Troubleshooting

Cookies help us deliver our services. By using our services, you agree to our use of cookies.