Fluorescence-Sensor Green
Description | Fluorescence-Sensor Green: excitation LED 525 nm, photodiode, Filter-Cap attached for Amplex UltraRed measurements. |
---|---|
Product ID | 44210-01 |
Type | O2k, MultiSensor, O2k-Fluorescence LED2-Module |
Link | O2k-Fluorescence LED2-Module |
Image |
O2k-Manual
- 2 Fluorescence-Sensors Green are included in the O2k-Fluorescence LED2-Module.
Mounting a Filter-Cap
The Filter-Cap of each sensor can be exchanged for application of different filter combinations on the same optical sensor.
- Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Insert a filter into either or both Filter-Cap windows for the LED or photodiode.
- Mounting: Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.
Connect
Connect the cable of the Fluorescence-Sensor to the Fluorescence-Sensor Plug of the Fluorescence-Control Unit. For recording of fluorescence signals, insert the black sensor into the window of the O2k chamber as far as possible, aligning the Sensor-Guide Sector with the Sensor-Guide of the O2k-Front Fixation.