Talk:O2k-Open Support alert: Difference between revisions
From Bioblast
No edit summary |
Garcia Luiz (talk | contribs) No edit summary |
||
Line 5: | Line 5: | ||
::::::: While running your experiment, especially during [[uncoupler titrations]], it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - Β»[[Smoothing#DatLab_oxygen_flux:_performance_and_data_analysis|Slope smoothing]]Β« | ::::::: While running your experiment, especially during [[uncoupler titrations]], it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - Β»[[Smoothing#DatLab_oxygen_flux:_performance_and_data_analysis|Slope smoothing]]Β« | ||
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] ''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Did you know that you can use the [[TIP2k-Module|Titration-Injection microPump (TIP2k)]] to perform an instrumental O<sub>2</sub> background on the [[O2k-sV-Module]]?''' | ||
::::::: By adjusting the [[Dithionite|dithionite]] concentration toΒ 2.5 mM, the [[MiPNet14.06 Instrumental O2 background|instrumental O<sub>2</sub> background]] can be performed with specific settings for the [[O2k-sV-Module]]. - Β»[[O2k-sV-Module]]Β« | ::::::: By adjusting the [[Dithionite|dithionite]] concentration toΒ 2.5 mM, the [[MiPNet14.06 Instrumental O2 background|instrumental O<sub>2</sub> background]] can be performed with specific settings for the [[O2k-sV-Module]]. - Β»[[O2k-sV-Module]]Β« |
Revision as of 10:05, 5 March 2020
- Do you perform the leaky chamber test after chamber assembly?
- After chamber reassembly you might experience a small leaky chamber only identified by a leaky chamber test- Β»Leaky chamber testΒ«
- Do you perform the leaky chamber test after chamber assembly?
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - Β»Slope smoothingΒ«
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with specific settings for the O2k-sV-Module. - Β»O2k-sV-ModuleΒ«
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?