|'''Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?''' <br>
|'''Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?''' <br>
- »[https://wiki.oroboros.at/index.php/Oxygen_flux_-_instrumental_background#Unstable_O2_slope_neg._signal_in_O2_instrumental_BG Instrumental background]« and »[[TIP2k Filter Papers]]«
- »[[Oxygen_flux_-_instrumental_background#Unstable_O2_slope_neg._signal_in_O2_instrumental_BG|Instrumental background]]« and »[[TIP2k Filter Papers]]«
|}
|}
:::: ''See the scheme'':
:::: ''See the scheme'':
<br>
<br>
[[Image:TEMPLATE ALERT.png|none|900px|Agenda/Alerts body text format]]
[[Image:TEMPLATE ALERT.png|none|900px|Agenda/Alerts body text format]]


== Bioblast alert 2020 ==
Prep for circular done here: [[Talk:Bioblast alert 2020]]


== MitoPedia alert ==
== MitoPedia alert ==
::: Example:
:{|
| width="110" align="center" | [[File:MitoPedia.jpg|left|80px|MitoPedia|link=MitoPedia]]
|''' Measuring fatty acid oxidation (FAO)-linked respiration requires a low concentration of malate '''<br>
In the absence of malate, accumulation of Acetyl-CoA inhibits FAO. <br>
- »'''[[Fatty acid oxidation pathway control state]]'''«, »'''[https://www.bioblast.at/index.php/Gnaiger_2020_BEC_MitoPathways Mitochondrial Pathways: Chapter 5]'''« <br>
- communicated by »[[Cecatto Cristiane]]« and »[[Cardoso Luiza]]«
|}


{|
| width="110" align="center" | [[File:MitoPedia.jpg|left|80px|MitoPedia]]
| '''Did you know that if [[ET_capacity|ET capacity]] (E) is lower than [[OXPHOS_capacity|OXPHOS capacity]] (''P'', [[ET_capacity#The_important_difference_between_states_P_and_E|''P''>''E'']]) in living cells or mitochondrial preparations, then you have encountered an experimental artefact? The possible triggers of this experimental artefact are:''' <br>
- loss of oxidative capacity during the time course of the respirometric assay,<br>
- insufficient [[Uncoupler|uncoupler]] concentrations,<br>
- over titration of an uncoupler, which inhibits ET capacity,<br>
- inhibition of ET capacity due to the use of high [[Oligomycin|oligomycin]] (inhibitor of the ATP synthase) concentrations employed to measure [[LEAK_respiration|LEAK respiration]] prior ET capacity.


Information from: Gnaiger Erich ''et al'' ― MitoEAGLE Task Group (2020) Mitochondrial physiology. Bioenerg Commun 2020.1 »[[BEC_2020.1_doi10.26124bec2020-0001.v1|Bioblast link]]«
=== Next ===
|}


[[User:Antunes Diana|Antunes Diana]] ([[User talk:Antunes Diana|talk]]) 11:13, 22 June 2020 (CEST)  Comments and suggestions relevant for checking:
::* For 'insufficient uncoupler concentrations' would not be more accurate 'insufficient uncoupler titrations' since we recommend always a titration approach with small volumes to avoid over titration? Maybe by writing 'insufficient uncoupler concentrations' it would lead the user towards using high U concentration per titration instead of using low volumes in each titration?
::* include relevant links (''e.g.,'' for ce and mtprep)?
::* Not sure if you agree in including the 'E' and 'P' icons within the alert - I thought it would be good just for a quick visualization of the terms.


Suggestion 2:
=== Under construction ===
{|
| width="110" align="center" | [[File:MitoPedia.jpg|left|80px|MitoPedia]]
| '''Did you know that if [[ET_capacity|ET capacity]] [[File:E.jpg]] is lower than [[OXPHOS_capacity|OXPHOS capacity]] [[File:P.jpg]] ( [[ET_capacity#The_important_difference_between_states_P_and_E|''P''>''E'']], [[ET_capacity#The_important_difference_between_states_P_and_E|''P''>''E'']]) in [[Living_cells | living cells]] or [[Mitochondrial_preparations | mitochondrial preparations]], then you have encountered an experimental artefact? The possible triggers of this experimental artefact are:''' <br>
- loss of oxidative capacity during the time course of the respirometric assay,<br>
- insufficient [[Uncoupler|uncoupler]] titrations,<br>
- over titration of an uncoupler, which inhibits ET capacity,<br>
- inhibition of ET capacity due to the use of high [[Oligomycin|oligomycin]] (inhibitor of the ATP synthase) concentrations employed to measure [[LEAK_respiration|LEAK respiration]] prior ET capacity.


Information from: Gnaiger Erich ''et al'' ― MitoEAGLE Task Group (2020) Mitochondrial physiology. Bioenerg Commun 2020.1 »[[BEC_2020.1_doi10.26124bec2020-0001.v1|Bioblast link]]«
on hold until the page is updated
:{|
| width="100" align="center" |[[File:FAT4BRAIN logo.png|right|80px|link=https://wiki.oroboros.at/index.php/FAT4BRAIN|FAT4BRAIN]]
|'''A FAO-mito-CNS platform to help fatty acid-related respiration measurements was developed in the FAT4BRAIN project''' <br>
- » [[FAT4BRAIN_FAO-mito-CNS_platform|FAT4BRAIN FAO-mito-CNS platform]] «
|}
 
== O2k-Brief alert ==
:::* Example:
:{|
| width="100" align="center" | [[File:O2k-brief.png|45px|link=http://wiki.oroboros.at/index.php/O2k-brief|O2k-brief repository]]
|'''Antitumoral properties of triterpenic acid functionalized gold nanoparticles arise from decreasing mitochondrial respiration''' <br>
* Mioc M, Mioc A, Racoviceanu R, Ghiulai R, Prodea A, Milan A, Barbu Tudoran L, Oprean C, Ivan V, Șoica C (2023) The antimelanoma biological assessment of triterpenic acid functionalized gold nanoparticles. [[Mioc 2023 Molecules|»Bioblast link«]]
|}
|}
=== Next ===
=== Next ===


== O2k-Publications alert ==
:::* Example:
:{|
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
|'''"Mitochondria–nucleus communication and its integration within the maintenance of cellular homeostasis constitute key determinants of cellular fate and organism health in the context of cellular stress, homeostasis perturbation, and disease"''' <br>
* Temelie M, Talpur R, Dominguez-Prieto M, Dantas Silva A, Cenusa C, Craciun L, Savu DI, Moisoi N (2023) Impaired integrated stress response and mitochondrial integrity modulate genotoxic stress impact and lower the threshold for immune signalling. [[Temelie 2023 Int J Mol Sci|»Bioblast link«]]
:from the O2k-Network »[[UK Leicester Moisoi N]]«
|}


===='''Author names of O2k-Publications/ Brief for the Agenda vs Bioblast''' ====


__NOTOC__ [[File:Expand.png|right|25px|Click to expand or collaps]]
<div class="toccolours mw-collapsible mw-collapsed">
'''Guidelines to update a reference'''
<div class="mw-collapsible-content">
::When creating an O2k-Publication/Brief alert, it is the responsibility of the author to check if the reference is accordingly to the correct guidelines:
[[File:Reference example.png|900px]]
::Correct format for a person called “First Name Surname”:
::In the gray box should be “Surname FN” - '''From where references to be used as O2k-Publication/Brief alerts are copied from.'''
[[File:Gray box reference.png|900px]]
::In the links should be “ Surname First N”
[[File:Authors names links reference.png|900px]]
::'''How to edit:'''
:::A) Check if all authors with profiles in Bioblast (blue links) have their profiles with full name. If they have, proceed to next step (B). If not, you need to move it following the steps “How to move a profile” below.
:::B) Click in “Edit with form”.
:::C) The gray box is edited in “Reference” and the links are changed in “Author(s)”. Change it accordingly and click in “save page” at the end of the page.
[[File:Editing reference.png|900px]]


'''How to move a profile (instructions sent by Marija):'''


* '''Jul-09 Th''' (to be confirmed)
So let’s take for example my page:  
:{|
[[Beno M]]  
| width="110" align="center" | [[File:MitoPedia.jpg|left|80px|MitoPedia]]
[https://wiki.oroboros.at/index.php?title=Beno_Marija&redirect=no]
| '''Do you know the different ways to adjust oxygen concentration in the Oroboros O2k?''' <br>
- »'''[[Setting_the_oxygen_concentration | Setting the oxygen concentration]]'''«
|}


== O2k-Brief alert ==
::As you can see there is a redirect from Beno Marija to Beno M. What has to be done is to switch the content of those pages. Here are the steps:
:::1. Go to e.g. “Beno M” on edit
:::2. Click on “Move”
:::3. Change the new title to the full name of the person. In my case it will be Beno M to Beno Marija
[[File:BenoMreference.jpg|500px]]
:::4. Now it should look like that:
[[File:BenoMarijaReference.png|500px]]
:::5. Click on “Move” and leave all the ticked boxes
:::6.Done
::After that you can continue in step B.
::If you have any doubt you can contact Cristiane, Marija or Mario.
</div>
</div>


=== Next ===
=== Next ===
April 25th


== O2k-Publications alert ==
:{|
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
|'''"Our results further enhance the understanding and relevance of bioenergetic function in liver ischemia and transplantation and provide the basis for further consideration of these parameters as biomarkers in LT [liver transplantation]."''' <br>
* Meszaros AT, Weissenbacher A, Schartner M, Egelseer-Bruendl T, Hermann M, Unterweger J, Mittelberger C, Reyer BA, Hofmann J, Zelger BG, Hautz T, Resch T, Margreiter C, Maglione M, Komlódi T, Ulmer H, Cardini B, Troppmair J, Öfner D, Gnaiger E, Schneeberger S, Oberhuber R (2024) The predictive value of graft viability and bioenergetics testing towards the outcome in liver transplantation.  [[Meszaros_2024_Transpl_Int|»Bioblast link«]]
:from the O2k-Network »[[AT Innsbruck Schneeberger S]] and [[AT Innsbruck Oroboros]]«
|}


=== Next ===
May 22nd


* '''Jun-30 Tu'''
:{|
:{|
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
|'''The antihyperglycemic Empagliflozin protects cardiac mitochondria against high-fat diet-induced dysfunction'''<br>
|'''"We have identified a circadian clock phase-dependent change in Complex I coupling efficiency, demonstrating for the first time a mitochondria–clock interaction in the hooded seal."''' <br>
:from the O2k-Network »[[LV Riga Makrecka-Kuka M]]«
* Ciccone C, Kante F, Folkow LP, Hazlerigg DG, West AC, Wood SH (2024) Circadian coupling of mitochondria in a deep-diving mammal. [[Ciccone 2024 J Exp Biol|»Bioblast link«]]
*Makrecka-Kuka M, Korzh S, Videja M, Vilks K, Cirule H, Kuka J, Dambrova M, Liepinsh E (2020) Empagliflozin protects cardiac mitochondrial fatty acid metabolism in a mouse model of diet-induced lipid overload. Cardiovasc Drugs Ther [Epub ahead of print]. [[Makrecka-Kuka 2020 Cardiovasc Drugs Ther|»Bioblast link«]]
:from the O2k-Network »[[NO Tromsoe Wood SH]]«
|}
|}


==  O2k-Open Support alert ==
==  MiPNet ==
:::* Example:
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' "No bubbles should be trapped in the electrolyte reservoir after membrane application and mounting"'''<br>
 
- »'''[[MiPNet28.10 SmartPOS-service]]'''«<br>
- communicated by »[[Timon-Gomez Alba]]« and »[[]]«
|}


=== Next ===
=== Next ===


::::* While in Home Office: [[User:Gnaigere|Gnaigere]] ([[User talk:Gnaigere|talk]]) 14:03, 19 March 2020 (CET) https://wiki.oroboros.at/index.php/Talk:O2k-Open_Support_alert  -  Many labs will move to home office mode. In line with our special DL7.4 home office offer (see Agenda today), therefore, I suggest to post in our Agenda '''analysis-linked''' O2k-Open Support cases during this time, and collect real-time operation cases for future releases.
May 2nd


* '''Jun-29 Mo'''
:{|
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]  
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''Find out how to evaluate the effect of the obtained ‘O2 slope neg’ noise on experimental results'''<br>
|''' "In respirometric studies on blood cells, the purity of cell preparation should always be emphasized and documented." '''<br>
- » [[Oxygen_sensor_test#Figure_2._POS_Quality_control_using_the_DatLab_protocol_.28MiPNet06.03_POS-calibration-SOP.29| QC1: Oxygen sensor test]]«
- »'''[[MiPNet21.17 BloodCellsIsolation]]'''«<br>
- communicated by »[[Timon-Gomez Alba]]« and »[[Baglivo Eleonora]]«
|}
|}


or
==  O2k-Open Support alert ==
 
:::*Examples:
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|25px|Info: take care]]
|'''Does your polarographic oxygen sensor (POS) need a deep clean?'''<br>
A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
<br>
- »'''[[MiPNet19.18B POS-service|POS service]]'''«<br>
- communicated by »[[Cecatto Cristiane]]« and »[[Leo Elettra]]«
|}
:{|
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]  
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]  
|'''How to decide whether the noise of the O2 slope neg. is within recommended values?'''<br>
|'''Did you know that a humid environment and/or liquid spillage may lead to a high raw POS signal [V]?'''<br>
- » [[Oxygen_calibration_-_DatLab#Instrumental_DL-Protocols| Noise of 'O2 slope neg.']]«
See how to clean and dry the electrical connections of the OroboPOS-Connector.<br>
- »'''[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]'''«<br>
- communicated by »[[Cecatto Cristiane]]« and »[[Baglivo Eleonora]]«
|}
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' Unusual stirrer bar rotation? '''<br>
A “jumping” stirrer or a slow rotation of the stirrer bar may indicate a problem either with the stirrer itself or with the cleaning of the glass chamber.
<br>
- »'''[https://wiki.oroboros.at/index.php/Stirrer-Bar%5Cwhite_PVDF%5C15x6_mm#Technical_support Stirrer bar - technical support]'''«<br>
- communicated by »[[Baglivo Eleonora]]« and »[[Grings Mateus]]«
|}
|}


(NOTE for 4-eyes: [[Oxygen_sensor_test#Figure_2._POS_Quality_control_using_the_DatLab_protocol_.28MiPNet06.03_POS-calibration-SOP.29| QC1: Oxygen sensor test]] is the O2k-Support page for sensor test, and [[Oxygen_calibration_-_DatLab#Instrumental_DL-Protocols| Noise of 'O2 slope neg.']] is the updated version by Erich that belongs to MitoPedia:_DatLab#Calibration. What to do with this pages? Shall the info remain in those locations or shall be linked instead?)
=== Next ===


=== Removed for correction ===


:{|
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''Do you perform the ''leaky chamber test'' after chamber assembly? '''<br>
|''' Do you observe an unstable O<sub>2</sub> slope negative following TIP2k titrations during an instrumental O<sub>2</sub> background test? '''<br>
An unnoticed leak from the chamber may affect your data, therefore we recommend performing the ''Leaky chamber test'' - especially after chamber reassembly. <br>
Incorrect placement of TIP2k Filter Papers can lead to liquid accumulation and convective oxygen backdiffusion. Check for proper placement by confirming they are fully in contact with the stopper receptacle.
- »[[Talk:O2k-Chamber_MitoPedia#Leaky_Chamber|Leaky chamber test]]« and » https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly«
<br>
- »'''[[MiPNet14.06_Instrumental_O2_background|Instrumental O<sub>2</sub> background]]'''«, »'''[https://wiki.oroboros.at/index.php/Oxygen_flux_-_instrumental_background#Unstable_O2_slope_neg._in_instrumental_background_test Unstable O<sub>2</sub> slope negative]'''«<br>
- communicated by »[[Grings Mateus]]« and »[[Baglivo Eleonora]]«
|}
|}


== O2k-Open Support alert circulars ==


=== O2k-Open Support alert circular 2020(01): in prep ===
April 15th
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''Do you observe either a zero or low raw O<sub>2</sub> signal when starting a measurement?'''<br>
Perform a test to evaluate whether this is related to an improperly positioned chamber or a faulty connection between the OroboPOS and the OroboPOS-Connector. <br>
- »'''[https://wiki.oroboros.at/index.php/Oxygen_signal#Raw_signal Oxygen signal]'''«<br>
- communicated by »[[Grings Mateus]]« and »[[Schmitt Sabine]]«
|}


April 29th
:{|
:{|
| width="110" align="center" | '''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.'''<br>
|'''How can I update DatLab 8 to the latest version?'''<br>
See how you can check that the two blue OroboPOS-Holders are intact. <br>
If your computer with DatLab 8 is connected to the internet, go to “Check for updates” in the Help menu of DatLab 8, and follow the simple installation instructions. <br>
- »[[O2k-OroboPOS-Holder_MitoPedia#O2k-OroboPOS-Holder:_Leaky_chamber_due_to_damaged_OroboPOS-Holder |O2k-OroboPOS-Holder]]«
- »'''[[Updating - DatLab]]'''«, »'''[https://www.oroboros.at/index.php/product/datlab/ DatLab 8]'''«<br>
- communicated by »[[Leo Elettra]]« and »[[Baglivo Eleonora]]«
|}
|}


== O2k-Feedback alert ==
:::* Example (if the person has a profile, link the name to the profile with [[Cecatto C| Cristiane Cecatto]]:
:{|
:{|
| width="110" align="center" | '''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|25px|Take care]]  
| width="110" align="center" | [[Image:O2k-Feedback.jpg|60px|link=https://wiki.oroboros.at/index.php/O2k-Feedback |O2k-Feedback]]
|'''Calibration parameters are transferred to experimental DatLab files'''<br>
|''' ''“Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”'''''<br>
Save calibration information from selected [[Marks - DatLab |Marks]] while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files.<br>  
Badr Ibrahim, PhD from Betterhumans Inc., US
- »[[Oxygen_calibration_-_DatLab#Oxygen_calibration_in_instrumental_and_experimental_DatLab_files | Oxygen calibration - DatLab]]«
<br>
- »'''[[O2k-Feedback]]'''«<br>
|}
|}
:{|
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]  
| width="110" align="center" | [[Image:O2k-Feedback.jpg|60px|link=https://wiki.oroboros.at/index.php/O2k-Feedback |O2k-Feedback]]
|'''Do you perform O2k-Chamber volume calibration?'''<br>  
|''' ''“Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”'''''<br>
The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the [[O2k-sV-Module]]). <br>
Badr Ibrahim, PhD from the O2k-Network »[[UK Leicester Moisoi N]]«<br>
- »[[O2k-Chamber MitoPedia]]«
- »'''[[O2k-Feedback]]'''«<br>
|}
|}


=== Next ===
== MitoFit Preprints ==
:::* Example:
:{|
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|25px|Take care]]  
| width="100" align="center"|[[Image:BEC-logo.png|100px|link=https://www.bioenergetics-communications.org/index.php/bec |Bioenergetics Communications]]
|'''Do you use the correct O<sub>2</sub> solubility factor (F<sub>M</sub>) in the ‘O2 calibration’ window in DatLab?<br>
|'''From preprint to BEC manuscript with published Open Peer Reviews:'''<br>
The O<sub>2</sub> solubility factor varies according to the medium used and influences the air calibration. <br>
*Komlódi T, Sobotka O, Gnaiger E (2021) Facts and artefacts on the oxygen dependence of hydrogen peroxide production using Amplex UltraRed. Bioenerg Commun 2021.4. [[doi:10.26124/BEC:2021-0004]]
- »[[Oxygen solubility factor]]«
:from the O2k-Network »[[AT Innsbruck Oroboros]]«  
|}
|}


== In preparation ==
=== Next ===
== For discussion ==
 
''Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020 ('''if''' the specific alert section is yet to be attributed)''
== Under discussion ==
 
This one needs an update of the Bioblast page and clarification of MiPNet06.03


:{|
:{|
| width="100" align="center" | [[Image:SUIT icon.jpg|80px|link=https://wiki.oroboros.at/index.php/MitoPedia:_SUIT |MitoPedia: SUIT]]  
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
| Controlled stability of the [https://www.oroboros.at/index.php/product/mitokit-cii MitoKit-CII] based on HPLC and LCMS analysis: No degradation of [[MitoKit-CII_MitoPedia| MitoKit-CII]] compounds [[MitoKit-CII/Succinate-nv| MitoKit-CII/Succinate-nv (NV118)]] and [[MitoKit-CII/Malonate-nv| MitoKit-CII/Malonate-nv (NV161)]] for following storage conditions:
|''' Did you enter the correct oxygen solubility factor in DatLab?'''<br>
* Proven -80 °C solid stability for more than 3 years (3 years 7 months)
The oxygen solubility factor of the incubation medium, F<sub>M</sub>, expresses the effect of the salt concentration on oxygen solubility relative to pure water. The correct oxygen solubility factor of your experimental medium or buffer needs to be entered in DatLab for proper calculation of the oxygen concentration.
* Proven -20 °C solid stability for 1 year
<br>
* Proven DMSO stability at room temperature for 1 week
- »'''[[Oxygen solubility factor|Oxygen solubility factor]]'''«<br>
* Proven DMSO stability at -20 °C for 6 months (ongoing)
- communicated by »[[Schmitt Sabine]]« and »[[ ]]«
 
|}
|}


== 2021 ==
This one needs a BioBlast link:


::::* 2021-01-06 https://www.jstor.org/stable/1706346?seq=1
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' When filling in the volume of uncoupler in the DatLab mark or the excel template, should one enter the total volume added or the volume to get to the highest flux?'''<br>
The volume added until the mark, equal to the highest flux, should be entered. Anything titrated after this mark should be summed to the other volumes before the next mark because this will be considered to calculate the sample dilution. 
<br>
- »'''[[LINK|LINK]]''' and '''[[MiPNet24.06_Oxygen_flux_analysis_-_DatLab_7.4|O<sub>2</sub> flux analysis with DatLab 7.4]]'''«<br>
- communicated by »[[Cardoso Luiza]]« and »[[ ]]«
|}


== Postponed events ==


:::: [[File:FAT4BRAIN logo.png|70px|FAT4BRAIN|link=FAT4BRAIN School Riga LV]] »<big>'''[[FAT4BRAIN School Riga LV | FAT4BRAIN School Riga LV IOC147]]'''</big
== Removed for correction ==
* '''Nov-30 Mo'''
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''Noisy oxygen signal? Check the OroboPOS and OroboPOS-Connector to clean electrical connections.'''<br>
Contaminating grease, salt crystals and moisture can be removed with a dry paper cloth or a paper cloth moistened with water and then absolute ethanol. <br>
- »[[OroboPOS-Connector#Cleaning_the_electrical_connections |Cleaning the electrical connections]]« <br>
- communicated by »[[Schmitt Sabine]]« and »[[Komlodi Timea]]«
|}
WGT recommended not to use water for cleaning. Erich introduced water for cleaning to remove salt crystals.


:::: »<big>'''[[APS2020 Chicago US]]'''</big>«


:::: [[Image:O2k-Network.png|70px|O2k-Workshops|link=O2k-Events#Next_O2k-Workshops]] [[MiPNet25.09 IOC San Diego CA US |<big>'''San Diego CA US'''</big>]]
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]
|'''Do you perform the ''leaky chamber test'' after chamber assembly? '''<br>
An unnoticed leak from the chamber may affect your data, therefore we recommend performing the ''Leaky chamber test'' - especially after chamber reassembly. <br>
- »[[Talk:O2k-Chamber_MitoPedia#Leaky_Chamber|Leaky chamber test]]« and » https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly«
|}

Latest revision as of 17:24, 23 April 2024

Antunes Diana (talk) 08:58, 5 May 2020 (CEST) Format for new Agenda/Alerts entries:
 {|
 | logos
 | title in bold <br>
 - link
 |}
 

e.g.,

O2k-Open SupportWarning: loss of data Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?

- »Instrumental background« and »TIP2k Filter Papers«

See the scheme:


Agenda/Alerts body text format


MitoPedia alert

Example:
MitoPedia
Measuring fatty acid oxidation (FAO)-linked respiration requires a low concentration of malate

In the absence of malate, accumulation of Acetyl-CoA inhibits FAO.
- »Fatty acid oxidation pathway control state«, »Mitochondrial Pathways: Chapter 5« 
- communicated by »Cecatto Cristiane« and »Cardoso Luiza« 


Next

Under construction

on hold until the page is updated
FAT4BRAIN
A FAO-mito-CNS platform to help fatty acid-related respiration measurements was developed in the FAT4BRAIN project

- » FAT4BRAIN FAO-mito-CNS platform «

O2k-Brief alert

  • Example:
O2k-brief repository Antitumoral properties of triterpenic acid functionalized gold nanoparticles arise from decreasing mitochondrial respiration
  • Mioc M, Mioc A, Racoviceanu R, Ghiulai R, Prodea A, Milan A, Barbu Tudoran L, Oprean C, Ivan V, Șoica C (2023) The antimelanoma biological assessment of triterpenic acid functionalized gold nanoparticles. »Bioblast link«

Next

O2k-Publications alert

  • Example:
O2k-Publications in the MiPMap "Mitochondria–nucleus communication and its integration within the maintenance of cellular homeostasis constitute key determinants of cellular fate and organism health in the context of cellular stress, homeostasis perturbation, and disease"
  • Temelie M, Talpur R, Dominguez-Prieto M, Dantas Silva A, Cenusa C, Craciun L, Savu DI, Moisoi N (2023) Impaired integrated stress response and mitochondrial integrity modulate genotoxic stress impact and lower the threshold for immune signalling. »Bioblast link«
from the O2k-Network »UK Leicester Moisoi N«

Author names of O2k-Publications/ Brief for the Agenda vs Bioblast

Click to expand or collaps

Guidelines to update a reference

When creating an O2k-Publication/Brief alert, it is the responsibility of the author to check if the reference is accordingly to the correct guidelines:

Reference example.png

Correct format for a person called “First Name Surname”:
In the gray box should be “Surname FN” - From where references to be used as O2k-Publication/Brief alerts are copied from.

Gray box reference.png

In the links should be “ Surname First N”

Authors names links reference.png

How to edit:
A) Check if all authors with profiles in Bioblast (blue links) have their profiles with full name. If they have, proceed to next step (B). If not, you need to move it following the steps “How to move a profile” below.
B) Click in “Edit with form”.
C) The gray box is edited in “Reference” and the links are changed in “Author(s)”. Change it accordingly and click in “save page” at the end of the page.

Editing reference.png

How to move a profile (instructions sent by Marija):

So let’s take for example my page: Beno M [1]

As you can see there is a redirect from Beno Marija to Beno M. What has to be done is to switch the content of those pages. Here are the steps:
1. Go to e.g. “Beno M” on edit
2. Click on “Move”
3. Change the new title to the full name of the person. In my case it will be Beno M to Beno Marija

BenoMreference.jpg

4. Now it should look like that:

BenoMarijaReference.png

5. Click on “Move” and leave all the ticked boxes
6.Done
After that you can continue in step B.
If you have any doubt you can contact Cristiane, Marija or Mario.

Next

April 25th

O2k-Publications in the MiPMap "Our results further enhance the understanding and relevance of bioenergetic function in liver ischemia and transplantation and provide the basis for further consideration of these parameters as biomarkers in LT [liver transplantation]."
  • Meszaros AT, Weissenbacher A, Schartner M, Egelseer-Bruendl T, Hermann M, Unterweger J, Mittelberger C, Reyer BA, Hofmann J, Zelger BG, Hautz T, Resch T, Margreiter C, Maglione M, Komlódi T, Ulmer H, Cardini B, Troppmair J, Öfner D, Gnaiger E, Schneeberger S, Oberhuber R (2024) The predictive value of graft viability and bioenergetics testing towards the outcome in liver transplantation. »Bioblast link«
from the O2k-Network »AT Innsbruck Schneeberger S and AT Innsbruck Oroboros«

May 22nd

O2k-Publications in the MiPMap "We have identified a circadian clock phase-dependent change in Complex I coupling efficiency, demonstrating for the first time a mitochondria–clock interaction in the hooded seal."
  • Ciccone C, Kante F, Folkow LP, Hazlerigg DG, West AC, Wood SH (2024) Circadian coupling of mitochondria in a deep-diving mammal. »Bioblast link«
from the O2k-Network »NO Tromsoe Wood SH«

MiPNet

  • Example:
O2k-Open SupportInfo: improve "No bubbles should be trapped in the electrolyte reservoir after membrane application and mounting"

- »MiPNet28.10 SmartPOS-service«
- communicated by »Timon-Gomez Alba« and »[[]]«

Next

May 2nd

O2k-Open SupportInfo: improve "In respirometric studies on blood cells, the purity of cell preparation should always be emphasized and documented."

- »MiPNet21.17 BloodCellsIsolation«
- communicated by »Timon-Gomez Alba« and »Baglivo Eleonora«

O2k-Open Support alert

  • Examples:
O2k-Open SupportInfo: take care Does your polarographic oxygen sensor (POS) need a deep clean?

A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
- »POS service«
- communicated by »Cecatto Cristiane« and »Leo Elettra«

O2k-Open SupportWarning: loss of data Did you know that a humid environment and/or liquid spillage may lead to a high raw POS signal [V]?

See how to clean and dry the electrical connections of the OroboPOS-Connector.
- »Cleaning the electrical connections«
- communicated by »Cecatto Cristiane« and »Baglivo Eleonora«

O2k-Open SupportInfo: improve Unusual stirrer bar rotation?

A “jumping” stirrer or a slow rotation of the stirrer bar may indicate a problem either with the stirrer itself or with the cleaning of the glass chamber.
- »Stirrer bar - technical support«
- communicated by »Baglivo Eleonora« and »Grings Mateus«

Next

O2k-Open SupportInfo: improve Do you observe an unstable O2 slope negative following TIP2k titrations during an instrumental O2 background test?

Incorrect placement of TIP2k Filter Papers can lead to liquid accumulation and convective oxygen backdiffusion. Check for proper placement by confirming they are fully in contact with the stopper receptacle.
- »Instrumental O2 background«, »Unstable O2 slope negative«
- communicated by »Grings Mateus« and »Baglivo Eleonora«


April 15th

O2k-Open SupportInfo: improve Do you observe either a zero or low raw O2 signal when starting a measurement?

Perform a test to evaluate whether this is related to an improperly positioned chamber or a faulty connection between the OroboPOS and the OroboPOS-Connector.
- »Oxygen signal«
- communicated by »Grings Mateus« and »Schmitt Sabine«

April 29th

O2k-Open SupportInfo: improve How can I update DatLab 8 to the latest version?

If your computer with DatLab 8 is connected to the internet, go to “Check for updates” in the Help menu of DatLab 8, and follow the simple installation instructions.
- »Updating - DatLab«, »DatLab 8«
- communicated by »Leo Elettra« and »Baglivo Eleonora«

O2k-Feedback alert

  • Example (if the person has a profile, link the name to the profile with Cristiane Cecatto:
O2k-Feedback “Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”

Badr Ibrahim, PhD from Betterhumans Inc., US
- »O2k-Feedback«

O2k-Feedback “Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”

Badr Ibrahim, PhD from the O2k-Network »UK Leicester Moisoi N«
- »O2k-Feedback«

Next

MitoFit Preprints

  • Example:
Bioenergetics Communications From preprint to BEC manuscript with published Open Peer Reviews:
  • Komlódi T, Sobotka O, Gnaiger E (2021) Facts and artefacts on the oxygen dependence of hydrogen peroxide production using Amplex UltraRed. Bioenerg Commun 2021.4. doi:10.26124/BEC:2021-0004
from the O2k-Network »AT Innsbruck Oroboros«

Next

Under discussion

This one needs an update of the Bioblast page and clarification of MiPNet06.03

O2k-Open SupportInfo: improve Did you enter the correct oxygen solubility factor in DatLab?

The oxygen solubility factor of the incubation medium, FM, expresses the effect of the salt concentration on oxygen solubility relative to pure water. The correct oxygen solubility factor of your experimental medium or buffer needs to be entered in DatLab for proper calculation of the oxygen concentration.
- »Oxygen solubility factor«
- communicated by »Schmitt Sabine« and »[[ ]]«

This one needs a BioBlast link:

O2k-Open SupportInfo: improve When filling in the volume of uncoupler in the DatLab mark or the excel template, should one enter the total volume added or the volume to get to the highest flux?

The volume added until the mark, equal to the highest flux, should be entered. Anything titrated after this mark should be summed to the other volumes before the next mark because this will be considered to calculate the sample dilution.
- »LINK and O2 flux analysis with DatLab 7.4«
- communicated by »Cardoso Luiza« and »[[ ]]«


Removed for correction

  • Nov-30 Mo
O2k-Open SupportInfo: improve Noisy oxygen signal? Check the OroboPOS and OroboPOS-Connector to clean electrical connections.

Contaminating grease, salt crystals and moisture can be removed with a dry paper cloth or a paper cloth moistened with water and then absolute ethanol.
- »Cleaning the electrical connections« 
- communicated by »Schmitt Sabine« and »Komlodi Timea«

WGT recommended not to use water for cleaning. Erich introduced water for cleaning to remove salt crystals.


O2k-Open SupportWarning: loss of data Do you perform the leaky chamber test after chamber assembly?

An unnoticed leak from the chamber may affect your data, therefore we recommend performing the Leaky chamber test - especially after chamber reassembly.
- »Leaky chamber test« and » https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly«