Difference between revisions of "Talk:Oxygen calibration - DatLab"

From Bioblast
Line 16: Line 16:


::; Indicator fields
::; Indicator fields
:::: As a reminder, the calibration indicator fields in the bottom right and bottom left corner of the experiment window are colored yellow after starting an experiment. After opening the calibration window by pressing F5 (or via the menu tool bar) and confirming the calibration, the field changes to green (in DatLab 8 as well as in DatLab 7.4).
:::: As a reminder, the calibration indicator fields in the bottom right and bottom left corner of the experiment window are colored yellow after starting an experiment. After opening the calibration window by pressing F5 (or via the top menu bar) and confirming the calibration, the field changes to green (in DatLab 8 as well as in DatLab 7.4).


; Zero oxygen calibration
; Zero oxygen calibration
:: Less frequent than air calibration, zero oxygen calibration should be done from time to time over weeks. It is mandatory after application of a new membrane and O2-sensor service to confirm sensor performance; ???bracketing zero oxygen calibrations when working at low oxygen???
:: Less frequent than air calibration, zero oxygen calibration should be done from time to time over weeks. In addition, zero oxygen calibration is recommended when the experimental workflow requires recording of low oxygen concentrations. It is mandatory after application of a new membrane and O2-sensor service to confirm sensor serviceability.


::; Instrumental O2 background test
::; Instrumental O2 background test

Revision as of 09:57, 18 October 2021

Bioblast wiki

Popular Bioblast page

Oxygen calibration - DatLab has been accessed more than
  • 5,000 times (2019-12-16)


                  


O2k-Open Support

Talk:Oxygen calibration - DatLab



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

O2 calibration is the calibration in DatLab of the oxygen sensor. It is a prerequisite for obtaining accurate measurements of respiration. Accurate calibration of the oxygen sensor depends on (1) equilibration of the incubation medium with air oxygen partial pressure at the temperature defined by the experimenter; (2) zero oxygen calibration; (3) high stability of the POS signal tested for sufficiently long periods of time; (4) linearity of signal output with oxygen pressure in the range between oxygen saturation and zero oxygen pressure; and (5) accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration. The standard oxygen calibration procedure is described below for high-resolution respirometry with the calibration routine using instrumental calibration DL-Protocols in DatLab.


Reference: MiPNet06.03 POS-calibration-SOP, MiPNet12.20_O2k-calibration_tutorial

Oxygen calibration - when and how?

Air calibration / Quality control
Daily air calibration goes hand in hand with a stirrer test and inspection of the recorded data in DatLab, also called oxygen sensor test. The procedure is decisive for maintaining a high level of O2k quality control as described in QC 1: Oxygen sensor test.
Indicator fields
As a reminder, the calibration indicator fields in the bottom right and bottom left corner of the experiment window are colored yellow after starting an experiment. After opening the calibration window by pressing F5 (or via the top menu bar) and confirming the calibration, the field changes to green (in DatLab 8 as well as in DatLab 7.4).
Zero oxygen calibration
Less frequent than air calibration, zero oxygen calibration should be done from time to time over weeks. In addition, zero oxygen calibration is recommended when the experimental workflow requires recording of low oxygen concentrations. It is mandatory after application of a new membrane and O2-sensor service to confirm sensor serviceability.
Instrumental O2 background test
As described in more detail in MiPNet14.06_Instrumental_O2_background, the instrumental O2 background test is necessary to assess the oxygen fluxes into or out of the respirometer chamber that are not directly related to respiratory activity of the sample. Is is recommended to perform it about once a month. Air calibration and zero oxygen calibration are part of the procedure.
Generate R1 and R0
  1. record the oxygen signal while the chamber is open
  2. after equilibration set a mark and calla it R1
    • or perform steps 1. and 2. guided by the air calibration protocol
  3. close the chamber, remove the oxygen from the medium
  4. set a mark and call it R0
    • or perform steps 3. and 4. guided by the air and zero calibration protocol

DatLab 8

  • How to get R1 and R0
  • calibration values are stored together with instrument identifiers and sensor numbers
  • bottom left and bottom right: yellow until calibration has been confirmed

DatLab 7.4

see current content of the page

Details

  • marks: what is it, how to set it, how to extract values
  • events: what is it, how to set it
  • special situation: one computer, several O2ks connected

Related Topics

  • Quality control --> link