:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how a switch components approach can address your faulty specifications even after countless quality control tests?''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how a switch components approach can address your faulty specifications even after countless quality control tests?''' | ||
::::::: Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform a experimental savior troubleshooting.-»[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]« | ::::::: Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform a experimental savior troubleshooting.-»[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]« | ||
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw signal [V]?''' | |||
::::::: See how to clean and dry the electrical connections of the OroboPOS-Connector.-»[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]« |
Revision as of 10:16, 10 March 2020
- Do you perform the leaky chamber test after chamber assembly?
- After chamber reassembly you might experience a small leaky chamber only identified by a leaky chamber test- »Leaky chamber test«
- Do you perform the leaky chamber test after chamber assembly?
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - »Slope smoothing«
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with specific settings for the O2k-sV-Module. - »O2k-sV-Module«
- Do you know how a switch components approach can address your faulty specifications even after countless quality control tests?
- Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform a experimental savior troubleshooting.-»Switch components approach«
- Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw signal [V]?
- See how to clean and dry the electrical connections of the OroboPOS-Connector.-»Cleaning the electrical connections«
- Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?