Leaky Chamber
Leaky chamber test
- When the leakage is so little that cannot be noticed in a short amount of time, we recommend to follow the Leaky chamber test below:
- Decrease the oxygen concentration down to ~ 50 ΞΌM (using Dithionite or N2 injection)
- Let the O2k record overnight with the chamber closed and illumination switched off
- The oxygen concentration should decrease overtime due to the O2 consumed by OroboPOS as it is shown in the figure. In this case, the chamber is assembled properly.
- If the oxygen concentration increases overtime, a reassemble of the chamber is needed.