Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "O2k-Barometric Pressure Transducer"

From Bioblast
Line 10: Line 10:
== Accuracy ==
== Accuracy ==


Quick accuracy check for the barometric pressure transducer 2013-10-21:
:::: Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):


* 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
::::* 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
* nearest metrological station ( Innsbruck airport, alt: 585 m above sea level): pressure = 95.5 kPa
::::* Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa
Β 
[[User:Fasching Mario|Fasching Mario]] 14:45, 21 October 2013 (CEST)

Revision as of 09:15, 12 March 2021

                



O2k-Barometric Pressure Transducer

O2k-Catalogue

Description O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years.
Product ID 20050
Type O2k-Main Unit
Link O2k-Core
Image
SI unit: kPa = JΒ·m-3
1 atm = 760 mmHg = 101.325 kPa

Accuracy

Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):
  • 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
  • Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa