Difference between revisions of "O2k-Barometric Pressure Transducer"
From Bioblast
Line 10: | Line 10: | ||
== Accuracy == | == Accuracy == | ||
Quick accuracy check for the barometric pressure transducer 2013-10-21: | :::: Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21): | ||
* 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa | ::::* 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa | ||
* | ::::* Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa | ||
Β | |||
Revision as of 09:15, 12 March 2021
O2k-Barometric Pressure Transducer
Description | O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years. |
---|---|
Product ID | 20050 |
Type | O2k-Main Unit |
Link | O2k-Core |
Image |
- SI unit: kPa = JΒ·m-3
- 1 atm = 760 mmHg = 101.325 kPa
Accuracy
- Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):
- 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
- Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa