|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration.
|title=[[Image:O2k-Manual.jpg|right|70px|link=O2k-Manual|O2k-Manual]]O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration.
|info=[[File:PDF.jpg|100px|link=http://wiki.oroboros.at/images/7/77/MiPNet06.03_POS-Calibration-SOP.pdf |Bioblast pdf]] |authors=OROBOROS
|info=[[File:PDF.jpg|100px|link=https://wiki.oroboros.at/images/7/77/MiPNet06.03_POS-Calibration-SOP.pdf |Bioblast pdf]] » [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions]
|year=2016-02-09
|authors=Oroboros
|year=2023-02-06
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract='''Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8.''' » [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions]
|abstract=Gnaiger E (2023) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(21):1-10.
 
{{MiPNet pdf page linking to MitoPedia}}
'''[[O2k-SOP |O2k-Protocols SOP]]''': [[High-resolution respirometry]] (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance ([[MiPNet19.18B POS-Service]]) and accurate calibration of the [[polarographic oxygen sensor]]s ([[OroboPOS]], POS). Standard operating procedures (O2k-SOP) are described as part of the [[MitoFit Quality Control System]]. (1) Clean and ready-to-use O2k-Chambers are a basic requirement ([[MiPNet19.03]]). (2) Quality control for evaluation of proper POS function (SOP: O2 sensor test). (3) Accurate POS calibration ([[MiPNet19.18D O2k-Calibration]]). This is part 1 of O2k Quality Control.
|mipnetlab=AT_Innsbruck_Oroboros
 
:» O2k-Protocols SOP: O2k Quality Control 2 »[[MiPNet14.06 InstrumentalO2Background]]«
:» O2k-Manual »[[MiPNet19.18D_O2k-Calibration]]«
:» Product: [[OroboPOS]], [[Oxygraph-2k|O2k]], [[OROBOROS O2k-Catalogue |O2k-Catalogue]]
:» Template: [[File:O2-calibration.xlsx|O2-calibration.xlsx]]
|mipnetlab=AT_Innsbruck_OROBOROS
}}
{{Labeling
|area=Instruments;methods
|instruments=Oxygraph-2k, O2k-Protocol
|additional=O2k-SOP
}}
}}


== Further details ==
== Keywords ==
{{Template:Keywords: Oxygen signal}}


[[File:O2k-QCS.jpg|350px|O2k-QCS]] Next step - O2k Quality Control 2: »[[MiPNet14.06 InstrumentalO2Background]]«
::::» [[Oxygen calibration - DatLab]]




[[Image:MiPNet06.03_Table.jpg|centre|frame|<sup>c</sup>O2* is the oxygen concentration at thermodynamic equilibrium with air under standard conditions of 100 kPa barometric pressure. ''F''<sub>M</sub> has been determined for MiR05 at 30 °C and 37 °C, and is assumed to be identical at 25 °C.]]
{{Labeling
 
|area=Respiration, Instruments;methods
== References ==
|instruments=Oxygraph-2k, O2k-Manual
::::* Forstner H, Gnaiger E (1983) Calculation of equilibrium oxygen concentration. In: Polarographic Oxygen Sensors. Aquatic and Physiological Applications. Gnaiger E, Forstner H (eds), Springer, Berlin, Heidelberg, New York:321-33. - [[Forstner_1983_POS |»Bioblast link«]]
|additional=DatLab, O2k-SOP, DL7, Oxygen solubility, DL8,  
::::* [[Gnaiger_2008_POS]]
MitoFit 2021 Dark respiration, MitoFit 2021 Photosynthesis, DL8, PLoSONE2022ace-sce, MitoFit2022QC, POS
::::* [[MiPNet19.18D_O2k-Calibration]]
}}
 
::: '''This article is cited in'''
::::* Rodriguez JV, Pizarro MD, Scandizzi AL, Guibert EE, Almada LL, Mamprin ME (2008) Construction and performance of a minibioreactor suitable as experimental bioartificial liver. Artif Organs 32:323-8. [http://www.ncbi.nlm.nih.gov/pubmed/18370948 »PubMed 18370948«]
 
 
[[Category:Technical service]]

Latest revision as of 14:27, 27 July 2023


                  


O2k-Open Support

MiPNet06.03 POS-calibration-SOP


Publications in the MiPMap
O2k-Manual
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration.

» Bioblast pdf  » Versions

Oroboros (2023-02-06) Mitochondr Physiol Network

Abstract: Gnaiger E (2023) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(21):1-10.

O2k-technical support and open innovation
Open the pdf document above.
» Current O2k-series: NextGen-O2k Series XB and O2k Series J
» Current software versions DatLab 8.0: MitoPedia: DatLab


O2k-Network Lab: AT_Innsbruck_Oroboros


Keywords


Questions.jpg


Click to expand or collaps


» Oxygen calibration - DatLab


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

DatLab, O2k-SOP, DL7, Oxygen solubility, DL8, MitoFit 2021 Dark respiration, MitoFit 2021 Photosynthesis, DL8, PLoSONE2022ace-sce, MitoFit2022QC, POS