Difference between revisions of "MiPNet06.03 POS-calibration-SOP"
Line 1: | Line 1: | ||
{{OROBOROS header page name}} | {{OROBOROS header page name}} | ||
{{Publication | {{Publication | ||
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] Polarographic oxygen sensors | |title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. | ||
|info=[[File:PDF.jpg|50px|link=http://wiki.oroboros.at/images/7/77/MiPNet06.03_POS-Calibration-SOP.pdf |Bioblast pdf]] » [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions] | |info=[[File:PDF.jpg|50px|link=http://wiki.oroboros.at/images/7/77/MiPNet06.03_POS-Calibration-SOP.pdf |Bioblast pdf]] » [http://www.bioblast.at/index.php/File:MiPNet06.03_POS-Calibration-SOP.pdf Versions] | ||
|authors=OROBOROS | |authors=OROBOROS | ||
|year=2016-02- | |year=2016-02-09 | ||
|journal=Mitochondr Physiol Network | |journal=Mitochondr Physiol Network | ||
|abstract=Gnaiger E (2016) Polarographic oxygen sensors | |abstract=Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8. | ||
[[High-resolution respirometry]] critically depends on accurate calibration of the [[polarographic oxygen sensor]]s | '''[[O2k-SOP |O2k-Protocols SOP]]''': [[High-resolution respirometry]] (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance ([[MiPNet19.18B POS-Service]]) and accurate calibration of the [[polarographic oxygen sensor]]s ([[OroboPOS]], POS). Standard operating procedures (O2k-SOP) are described as part of the [[MitoFit Quality Control System]]. (1) Clean and ready-to-use O2k-Chambers are a basic requirement ([[MiPNet19.03]]). (2) Quality control for evaluation of proper POS function (SOP: O2 sensor test). (3) Accurate POS calibration ([[MiPNet19.18D O2k-Calibration]]). This is part 1 of O2k Quality Control. | ||
:» O2k-Manual | |||
:» O2k-Protocols SOP: O2k Quality Control 2 »[[MiPNet14.06 InstrumentalO2Background]]« | |||
:» O2k-Manual »[[MiPNet19.18D_O2k-Calibration]]« | |||
:» Product: [[OroboPOS]], [[Oxygraph-2k|O2k]], [[OROBOROS O2k-Catalogue |O2k-Catalogue]] | :» Product: [[OroboPOS]], [[Oxygraph-2k|O2k]], [[OROBOROS O2k-Catalogue |O2k-Catalogue]] | ||
:» Template: [[File:O2k-Calibration-List.xls|O2k Calibration List]] | :» Template: [[File:O2k-Calibration-List.xls|O2k Calibration List]] | ||
Line 19: | Line 21: | ||
|additional=O2k-SOP | |additional=O2k-SOP | ||
}} | }} | ||
__TOC__ | |||
{{Technical support integrated}} | {{Technical support integrated}} | ||
[[File:O2 sensor test.jpg|right|400px|link=O2 sensor test |O2 sensor test]] | [[File:O2 sensor test.jpg|right|400px|link=O2 sensor test |O2 sensor test]] | ||
== Further details == | |||
* Forstner H, Gnaiger E (1983) Calculation of equilibrium oxygen concentration. In: Polarographic Oxygen Sensors. Aquatic and Physiological Applications. Gnaiger E, Forstner H (eds), Springer, Berlin, Heidelberg, New York:321-33. [[Forstner_1983_POS |»Bioblast Access«]] | * Forstner H, Gnaiger E (1983) Calculation of equilibrium oxygen concentration. In: Polarographic Oxygen Sensors. Aquatic and Physiological Applications. Gnaiger E, Forstner H (eds), Springer, Berlin, Heidelberg, New York:321-33. [[Forstner_1983_POS |»Bioblast Access«]] | ||
[[File:O2k-QCS.jpg|350px|O2k-QCS]] Next step - O2k Quality Control 2: »[[MiPNet14.06 InstrumentalO2Background]]« | |||
Revision as of 08:28, 9 February 2016
MiPNet06.03 POS-calibration-SOP
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. |
» » Versions
OROBOROS (2016-02-09) Mitochondr Physiol Network
Abstract: Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8.
O2k-Protocols SOP: High-resolution respirometry (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance (MiPNet19.18B POS-Service) and accurate calibration of the polarographic oxygen sensors (OroboPOS, POS). Standard operating procedures (O2k-SOP) are described as part of the MitoFit Quality Control System. (1) Clean and ready-to-use O2k-Chambers are a basic requirement (MiPNet19.03). (2) Quality control for evaluation of proper POS function (SOP: O2 sensor test). (3) Accurate POS calibration (MiPNet19.18D O2k-Calibration). This is part 1 of O2k Quality Control.
- » O2k-Protocols SOP: O2k Quality Control 2 »MiPNet14.06 InstrumentalO2Background«
- » O2k-Manual »MiPNet19.18D_O2k-Calibration«
- » Product: OroboPOS, O2k, O2k-Catalogue
- » Template: O2k Calibration List
• O2k-Network Lab: AT_Innsbruck_OROBOROS
Labels: MiParea: Instruments;methods
HRR: Oxygraph-2k, O2k-Protocol
O2k-SOP
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Further details
- Forstner H, Gnaiger E (1983) Calculation of equilibrium oxygen concentration. In: Polarographic Oxygen Sensors. Aquatic and Physiological Applications. Gnaiger E, Forstner H (eds), Springer, Berlin, Heidelberg, New York:321-33. »Bioblast Access«
Next step - O2k Quality Control 2: »MiPNet14.06 InstrumentalO2Background«
References
This article is cited in
- Rodriguez JV, Pizarro MD, Scandizzi AL, Guibert EE, Almada LL, Mamprin ME (2008) Construction and performance of a minibioreactor suitable as experimental bioartificial liver. Artif Organs 32:323-8. »PubMed 18370948«