Difference between revisions of "High signal at zero oxygen"
From Bioblast
Line 2: | Line 2: | ||
__TOC__ | __TOC__ | ||
{{MitoPedia without banner | {{MitoPedia without banner | ||
|abbr=''R''0 | |||
|description='A ''high signal at zero oxygen''' may be observed during [[zero calibration]]. The following instructions show how to further localize and possibly solve this technical problem. | |description='A ''high signal at zero oxygen''' may be observed during [[zero calibration]]. The following instructions show how to further localize and possibly solve this technical problem. | ||
|info= [[Zero_calibration | Zero calibration]], [[OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections | OroboPOS-Connector]] | |||
}} | }} | ||
== High zero current == | == High zero current == |
Revision as of 17:08, 29 October 2020
High signal at zero oxygen |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
'A high signal at zero oxygen' may be observed during zero calibration. The following instructions show how to further localize and possibly solve this technical problem.
Abbreviation: R0
Reference: Zero calibration, OroboPOS-Connector
High zero current
- In order to localize the problem, the following test should be performed:
-
- Note: Before sending us the O2k or any spare components, please contact us via our O2k-Open Support system: O2k-Open Support
-
- Example:
- Figure 1. High zero current test. Upper panel: Zero current is in the range of ±0.00099 V. Lower panel: Zero current is higher than 0.00099 V when the OroboPOS-Connector is connected to the O2k-Main unit and when the OroboPOS is connected to the OroboPOS-Connector, which means that both the OroboPOS-Connector and OroboPOS have electrical problems. First, the OroboPOS-Connector and the POS were disassembled and disconnected from the main unit. Second, the OroboPOS-Connector was connected to the O2k-Main unit without the OroboPOS. Third, the OroboPOS was connected to the OroboPOS-Connector without membranes. After each step, the O2 raw signal [V] was recorded for 10 min. The marks were set to the O2 raw signal [V] to evaluate the zero current.
- Example:
Keywords
- Bioblast links: Oxygen signal - >>>>>>> - Click on [Expand] or [Collapse] - >>>>>>>
- Specific
- O2k-Procedures
- MiPNet O2k-Procedures
- Other keyword lists
MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry