Difference between revisions of "Air calibration"

From Bioblast
 
(16 intermediate revisions by 4 users not shown)
Line 1: Line 1:
'''Air calibration''' of the polarographic oxygen sensor is performed routinely on any day before an experiment. Switch on the O2k, select the experimental temperature and gain setting (Oxygraph control window [F7]). Clean the chamber and add experimental medium (at least 2.1 ml for a 2 ml chamber). Then the stopper is fully inserted while the stirrer is on, and excess medium is siphoned off the receptacle of the stopper. Now the stopper is partially opened and positiond with the [[stopper spacer]].  
{{MitoPedia
|abbr=R1
|description='''Air calibration''' of an oxygen sensor (polarographic oxygen sensor) is performed routinely on any day before starting a respirometric experiment. The volume fraction of oxygen in dry air is constant. An aqueous solution in equilibrium with air has the same partial pressure as that in water vapour saturated air. The water vapour is a function of temperature only. The partial oxygen pressure in aqueous solution in equilibrium with air is, therefore, a function of total barometric pressure and temperature. Bubbling an aqueous solution with air generates deviations from barometric pressure within small gas bubbles and is, therefore, not recommended. To equilibrate an aqueous solution ata known partial pressure of oxygen [kPa], the aqueous solution is stirred rigorously in a chamber enclosing air at constant temperature. The concentration of oxygen, ''c''<sub>O2</sub> [µM], is obtained at any partial pressure by multiplying the partial pressure by the oxygen solubility, ''S''<sub>O2</sub> [µM/kPa]. ''S''<sub>O2</sub> is a function of temperature and composition of the salt solution, and is thus a function of the experimental medium. The [[Oxygen_solubility_factor|solubility factor]] of the medium, ''F''<sub>M</sub>, expresses the oxygen solubility relative to pure water at any experimental temperature. ''F''<sub>M</sub> is 0.89 in serum (37 °C) and 0.92 in [[MiR06]] or [[MiR05]] (30 °C and 37 °C).
|info=[[MiPNet06.03 POS-calibration-SOP]], [[Oxygen calibration - DatLab]], [[Oxygen sensor test]]
}}


Then the stirred aqeous phase is equilibrated with the oxygen contained in the small air space included in the chamber. Use [[Graph layout]] "01 Calibration Exp. Gr3-Temp" or "02 Background Experiment".
{{Template:Keywords: Oxygen signal}}


After stability is obtained within 20-30 min (the [[uncorrected slope]] is close to zero (less than +/- 1 pmol.s<sup>-1</sup>.ml<sup>-1</sup>), a mark is set on the blue trace of the oxygen signal (R1), and inserted in the calibration window [F5].
{{MitoPedia methods
 
|mitopedia method=Respirometry
Perform the oxygen calibration [F5] on-line, such that the calibration parameters are carried over automatically as a default calibration in subsequently opened DatLab files.
}}
 
{{MitoPedia O2k and high-resolution respirometry
Calibration parameters are copied to clipboard, for insertion into the Excel Template "[[O2k-Calibration-List]].xls". This template can be copied from [http://www.oroboros.at/index.php?o2k-o2calibration www.oroboros.at].
|mitopedia O2k and high-resolution respirometry=DatLab, Oroboros QM
 
}}
See also: [[Raw signal]], [[Zero calibration]], [[Stirrer test]]
{{MitoPedia topics
 
|mitopedia topic=Medium
* [http://www.oroboros.at/index.php?o2k-o2calibration O2k-Manual: Oxygen and pX (pH) calibration. MiPNet12.08]
}}
* [[MiPNet06.03_O2-Calib-Solubility]]
* [[Gnaiger_2008_POS]]
 
{{#set:Technical service=POS calibration}}
 
__SHOWFACTBOX__
 
{{Technical service}}

Latest revision as of 10:01, 5 October 2020


high-resolution terminology - matching measurements at high-resolution


Air calibration

Description

Air calibration of an oxygen sensor (polarographic oxygen sensor) is performed routinely on any day before starting a respirometric experiment. The volume fraction of oxygen in dry air is constant. An aqueous solution in equilibrium with air has the same partial pressure as that in water vapour saturated air. The water vapour is a function of temperature only. The partial oxygen pressure in aqueous solution in equilibrium with air is, therefore, a function of total barometric pressure and temperature. Bubbling an aqueous solution with air generates deviations from barometric pressure within small gas bubbles and is, therefore, not recommended. To equilibrate an aqueous solution ata known partial pressure of oxygen [kPa], the aqueous solution is stirred rigorously in a chamber enclosing air at constant temperature. The concentration of oxygen, cO2 [µM], is obtained at any partial pressure by multiplying the partial pressure by the oxygen solubility, SO2 [µM/kPa]. SO2 is a function of temperature and composition of the salt solution, and is thus a function of the experimental medium. The solubility factor of the medium, FM, expresses the oxygen solubility relative to pure water at any experimental temperature. FM is 0.89 in serum (37 °C) and 0.92 in MiR06 or MiR05 (30 °C and 37 °C).

Abbreviation: R1

Reference: MiPNet06.03 POS-calibration-SOP, Oxygen calibration - DatLab, Oxygen sensor test



Questions.jpg


Click to expand or collaps



MitoPedia methods: Respirometry 


MitoPedia O2k and high-resolution respirometry: DatLab, Oroboros QM 


MitoPedia topics: Media for respirometry