{{Technical support}} __TOC__ {{MitoPedia without banner |abbr= |description=[[Image:Stopper black PEEK conical Shaft central Port.JPG|right|180px]]'''Stopper\black PEEK\conical Shaft\central Port''': with conical shaft (with PTFE, graphite, carbon fiber) and one central capillary (1.3 mm diameter; 50.6 mm length), [[Volume-Calibration Ring]] (A or B) for volume adjustment 1.5 to 3.2 ml; 2 mounted O-rings, with 8 spare O-rings ([[O-ring\Viton\12.5x1 mm]]). The black PEEK stoppers are required for optical O2k-MultiSensor Modules. |info=[https://www.oroboros.at/index.php/product/stopperblack-peekconical-shaftcentral-port/ '''Product details and purchase information'''] }} == O2k-Manual == === Stoppers and optical measurements === :::: Use only black PEEK stoppers in conjunction with optical measurements. If necessary, replace the previous [[Stopper\white PVDF\conical Shaft\central Port |white PVDF stoppers]]. The black stoppers can be used for all HRR applications in general. See [[MiPNet22.11 O2k-FluoRespirometer manual]] for calibration of the 2-mL [[O2k-chamber]] volume, which is identical for PEEK and PVDF stoppers. == Keywords == {{Template:Keywords: Chamber volume}}